FDA Adverse Event Death Summary report: N

TABLO HEMODIALYSIS SYSTEM

MDR report key: 11009462 · Received December 15, 2020

Report

Report Number
3010355846-2020-00033
Event Type
Death
Date Received
December 15, 2020
Date of Event
November 14, 2020
Report Date
January 13, 2021
Manufacturer
OUTSET MEDICAL, INC.
Product Code
KDI
UDI-DI
00850001011112
PMA / PMN Number
K190793
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

PLEASE NOTE THAT THE FOLLOWING SECTION WAS UPDATED WITH ADDITIONAL INFORMATION RECEIVED ON 15-DEC-2020: B7: OTHER RELEVANT HISTORY, INCLUDING PREEXISTING MEDICAL CONDITIONS. G3/4: DATE RECEIVED BY MANUFACTURER.

Additional Manufacturer Narrative · 1

FROM THE INFORMATION PROVIDED, THERE IS NO INDICATION THAT THERE WAS ANY DEVICE MALFUNCTION, NONCONFORMANCE, OR MISUSE THAT CONTRIBUTED TO THE REPORTED EVENT. POTENTIAL ADVERSE EVENTS IN THE INSTRUCTIONS FOR USE (IFU) WITH THE TABLO SYSTEM INCLUDES, BUT ARE NOT LIMITED TO, OTHER, MORE SERIOUS, COMPLICATIONS ARISING FROM DIALYSIS, SUCH AS HEMORRHAGE, AIR EMBOLISM, ACIDOSIS, ALKALOSIS OR HEMOLYSIS, CAN CAUSE SERIOUS PATIENT INJURY OR DEATH. OUTSET MEDICAL, INC. TECHNICAL SUPPORT SPECIALIST (TSS) HAS REVIEWED SITE SYSTEM LOGS WITH A PROCEDURE DATE OF (B)(6) 2020. NO RELATED SYSTEM ALARMS WERE FOUND TO HAVE OCCURRED DURING TREATMENT. THE DEVICE IS FUNCTIONING POST TREATMENT. A REVIEW OF PRODUCTION RECORDS FOR THIS UNIT DID NOT NOTE ANY MANUFACTURING NONCONFORMANCES THAT WOULD CONTRIBUTE TO A PRODUCT EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A 12 HOUR DIALYSIS TREATMENT, THE PATIENT PASSED AWAY. IT WAS REPORTED THAT THE PATIENT HAD MULTIPLE UNSPECIFIED MEDICAL PROBLEMS AND WAS IN BAD HEALTH CONDITIONS BEFORE STARTING DIALYSIS TREATMENT ON TABLO. PER THE INFORMATION RECEIVED FROM THE CUSTOMER, THIS EVENT IS ATTRIBUTED TO THE PATIENT'S PRE-EXISTING CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1476239 TABLO HEMODIALYSIS SYSTEM DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI OUTSET MEDICAL, INC. TABLO X HEMODIALYSIS SYSTEM 00850001011112

Patients

Seq Age Sex Outcome Treatment
1 Death