FDA Adverse Event Other Summary report: N

9610617-2008-00010

MDR report key: 1100909 · Received May 16, 2008

Report

Report Number
9610617-2008-00010
Event Type
Other
Date Received
May 16, 2008
Product Code
HRX
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

THE 28305BA/SCOPE IS A 1.9MM X 6.5CM 30 DEGREE SCOPE. THE SHAFT OF THE SCOPE IS BENT; THERE ARE 2+ BROKEN ROD LENSES; THERE ARE NO VISUALS. THE 28306BN SHEATH IT WAS USED WITH WAS BENT AND DENTED. THERE WAS CONSIDERABLE RESISTANCE FELT WHEN WE INTRODUCED SCOPE INTO SHEATH. WE BELIEVE THE INTERFACE BETWEEN THE SCOPE AND THE BENT SHEATH CAUSED THE SCOPE SHAFT TO BE BENT RESULTING IN BROKEN ROD LENSES AND NO VISUALS. THERE HAVE BEEN NO COMPLAINTS RECORDED ON THIS SCOPE FOR THE LAST 6 YRS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HRX

Patients

Seq Age Sex Outcome Treatment
1