FDA Adverse Event Malfunction Summary report: N

INTIMA-II Y18GAX1.16IN PRN/EC SLM

MDR report key: 11008614 · Received December 15, 2020

Report

Report Number
3006948883-2020-00948
Event Type
Malfunction
Date Received
December 15, 2020
Date of Event
November 7, 2020
Report Date
January 6, 2021
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 0140311. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT INTIMA-II Y18GAX1.16IN PRN/EC SLM CATHETER BROKE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CATHETER TUBE WAS BROKEN WHEN THE NEEDLE WAS PIERCED INTO THE BLOOD VESSEL AND PERFORMING PUNCTURING.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT INTIMA-II Y18GAX1.16IN PRN/EC SLM CATHETER BROKE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CATHETER TUBE WAS BROKEN WHEN THE NEEDLE WAS PIERCED INTO THE BLOOD VESSEL AND PERFORMING PUNCTURING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1479251 INTIMA-II Y18GAX1.16IN PRN/EC SLM CATHETER FOZ BD (SUZHOU) 0140311

Patients

Seq Age Sex Outcome Treatment
1