FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 11008467 · Received December 15, 2020

Report

Report Number
2024168-2020-10480
Event Type
Injury
Date Received
December 15, 2020
Date of Event
November 7, 2020
Report Date
March 3, 2021
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED ¿SUTURE DID NOT LEAVE THE DEVICE¿ WAS CONFIRMED AS A CUFF MISS. A REVIEW OF THE MANUFACTURING RECORDS IDENTIFIED NO MANUFACTURING NONCONFORMITIES. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.D4 LOT NUMBER UPDATED FROM 9111341 TO 9111441.

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE VIA A 14F SHEATH RELATIVE TO A TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) PROCEDURE. REPORTEDLY, THE PROGLIDE SUTURE DID NOT LEAVE THE DEVICE. ANOTHER PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1478599 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 9111441

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 14F| SHEATH: 14F