AMIA AUTOMATED PD CYCLER SET
Report
- Report Number
- 1416980-2020-07765
- Event Type
- Malfunction
- Date Received
- December 15, 2020
- Date of Event
- November 19, 2020
- Report Date
- January 19, 2021
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FKX
- PMA / PMN Number
- K151525
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
BRAND NAME IS AMIA AUTOMATED PD CYCLER SET. CATALOGUE # IS 5C5479. DEVICE MANUFACTURER NAME: BAXTER HEALTHCARE - MOUNTAIN HOME. PMA/510K IS K151525. ONE ACTUAL SAMPLE WAS RETURNED FOR EVALUATION WITH THE PATIENT CONNECTOR ATTACHED TO THE FEMALE CONNECTOR OF A TRANSFER SET. THE FEMALE CONNECTOR WAS CONNECTED, DISCONNECTED, AND LEAK TESTED WITH THE AMIA CONNECTOR WITH NO ISSUES NOTED. THE REPORTED CONDITION WAS NOT VERIFIED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE DEVICE WAS RECEIVED AND IS CURRENTLY AWAITING EVALUATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THERE WAS DIFFICULTY DISCONNECTING THE FEMALE CONNECTOR OF THE TRANSFER SET FROM THE PATIENT LINE OF THE HOMECHOICE CASSETTE. IT WAS FURTHER REPORTED THAT THE DARK BLUE END INSIDE THE TRANSFER SET IS UNGLUED AND COMING OUT OF THE TRANSFER SET AND THE POINTED END REMAINS IN THE PATIENT CONNECTION LINE. THIS OCCURRED DURING AN UNSPECIFIED PROCESS STEP OF PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1471926 | AMIA AUTOMATED PD CYCLER SET | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE CORPORATION | NA | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | TRANSFER SET |