LINER NEUTRAL 36 MM I.D. SIZE D
Report
- Report Number
- 0001822565-2020-04049
- Event Type
- Injury
- Date Received
- December 14, 2020
- Date of Event
- November 24, 2020
- Report Date
- December 17, 2020
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- LPH
- UDI-DI
- 00889024519893
- PMA / PMN Number
- K190660
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE REPORTED EVENT OF DEEP INFECTION <90 DAYS OCCURRED POST IMPLANTATION. DURING THE INVESTIGATION PROCESS A REVIEW OF THE STERILE CERTIFICATIONS WERE REVIEWED AND FOUND TO BE CONFORMING WITH NO APPLICABLE DEVIATIONS. DEVICES WERE VERIFIED TO HAVE GONE THROUGH ACCEPTABLE STERILIZATION PROCESS FOLLOWING ISO/AAMI/ASTM & EU PUBLISHED GUIDELINES. THERE ARE MULTIPLE FACTORS THAT MAY CONTRIBUTE TO AN INFECTION THAT ARE OUTSIDE THE CONTROL OF ZIMMER BIOMET, SUCH AS EXTERNAL FACTORS, I.E. HOSPITAL/SURGICAL ENVIRONMENT, PROVIDER RELATED RISK FACTORS, AND/OR PATIENT COMORBIDITIES/RISK FACTORS. AS THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUES IDENTIFIED AFFECTING IMPLANT SAFETY OR EFFECTIVENESS, THEREFORE IMPLANTED PRODUCTS ARE NOT IDENTIFIED AS THE SOURCE OR CONTRIBUTING TO THE REPORTED INFECTION.
THE REPORTED EVENT OF DEEP INFECTION <90 DAYS OCCURRED POST IMPLANTATION. DURING THE INVESTIGATION PROCESS A REVIEW OF THE STERILE CERTIFICATIONS WERE REVIEWED AND FOUND TO BE CONFORMING WITH NO APPLICABLE DEVIATIONS. DEVICES WERE VERIFIED TO HAVE GONE THROUGH ACCEPTABLE STERILIZATION PROCESS FOLLOWING ISO/AAMI/ASTM & EU PUBLISHED GUIDELINES. THERE ARE MULTIPLE FACTORS THAT MAY CONTRIBUTE TO AN INFECTION THAT ARE OUTSIDE THE CONTROL OF ZIMMER BIOMET, SUCH AS EXTERNAL FACTORS, I.E. HOSPITAL/SURGICAL ENVIRONMENT, PROVIDER RELATED RISK FACTORS, AND/OR PATIENT COMORBIDITIES/RISK FACTORS. AS THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUES IDENTIFIED AFFECTING IMPLANT SAFETY OR EFFECTIVENESS, THEREFORE IMPLANTED PRODUCTS ARE NOT IDENTIFIED AS THE SOURCE OR CONTRIBUTING TO THE REPORTED INFECTION.
(B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, AS THE DEVICE WAS REQUESTED BUT NOT RETURNED BY THE HOSPITAL. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: PART: 650-0662, LOT: 3030964.
IT WAS REPORTED THAT PATIENT UNDERWENT A REVISION PROCEDURE TWO MONTHS POST IMPLANTATION DUE TO A SUSPECTED ACUTE INFECTION. THE SURGEON PERFORMED A WASHOUT WITH A HEAD A POLY SWAP. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1467821 | LINER NEUTRAL 36 MM I.D. SIZE D | PROTHESIS, HIP | LPH | ZIMMER BIOMET, INC. | N/A | 64498880 | 00889024519893 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Hospitalization| R | SEE H10 |