FDA Adverse Event Malfunction Summary report: N

EVIS EXERA II DUODENOVIDEOSCOPE

MDR report key: 11001598 · Received December 14, 2020

Report

Report Number
8010047-2020-10316
Event Type
Malfunction
Date Received
December 14, 2020
Report Date
January 18, 2021
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDT
UDI-DI
04953170229503
PMA / PMN Number
K143153
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE SUBJECT DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED PHENOMENON COULD NOT BE CONCLUSIVELY DETERMINED.

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC BUT WAS RETURNED TO OLYMPUS (B)(4). (B)(4) SENT THE DEVICE TO A THIRD PARTY LABORATORY FOR MICROBIOLOGICAL TESTING. AS A RESULT OF THE TESTING, THE SAMPLE COLLECTED FROM THE INSTRUMENT CHANNEL OF THE DEVICE TESTED POSITIVE FOR UNSPECIFIED MICROBES (10 CFU/CHANNEL) AND THE AIR/WATER CHANNEL OF THE DEVICE TESTED POSITIVE FOR UNSPECIFIED MICROBES (7 CFU/CHANNEL) . THE TESTING RESULT CLEARED THE GERMAN GUIDELINE. ODE CHECKED THE SUBJECT DEVICE AND FOUND FOLLOWING. THE BENDING SECTION RUBBER ADHESIVE WAS DAMAGED. THE ELECTRICAL CONNECTOR HAD LEAKAGE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF ROUTINE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. UNSPECIFIED MICROBES (2 CFU), PANTOEA SEPTICA (8 CFU), MORAXELLA OSLOENSIS (6 CFU) THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, INNOVA E4, USING ALDEHYDE. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1470551 EVIS EXERA II DUODENOVIDEOSCOPE DUODENOVIDEOSCOPE FDT OLYMPUS MEDICAL SYSTEMS CORP. TJF-Q180V 04953170229503

Patients

Seq Age Sex Outcome Treatment
1