ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAFETY LOCK
Report
- Report Number
- 1527736-1997-01863
- Event Type
- Malfunction
- Date Received
- July 31, 1997
- Report Date
- July 6, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- NURSE
Narratives
FACILITY EXPERIENCED AN EVENT WITH ENDOPATH* THORACIC ENDOSCOPIC LINEAR CUTTER WITH SAFETY LOCK ON WHILE PERFORMING A THORACOSCOPY. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #974254. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: LOCKOUT POSITION, UNFIRED; CARTRIDGE CONDITIONS, UNFIRED; CARTRIDGE RETURN BATCH NUMBER, J00M6X AND INSTRUMENT NUMBER, 98. FUNCTIONAL TESTS & RESULTS: CONDITION OF FIRING TRIGGER LOCKOUT, GOOD; CONDITION OF PINION GEAR, GOOD; CONDITION OF SHORT RACK, GOOD; CONDITION OF YOKE, GOOD AND WAS INSTRUMENT CYCLED YES. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO RECEIVED, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHY THE INSTRUMENT REPORTEDLY "DROPPED THE RELOAD" DURING SURGERY. THE INSTRUMENT WAS RETURNED IN GOOD PHYSICAL CONDITION. THE CARTRIDGE RETENTION FEATURE WAS MEASURED AND WAS FOUND TO BE WITHIN DESIGN SPECIFICATION. THE INSTRUMENT WAS CYCLED, FIRED, CUT, AND FORMED THE STAPLES WITHIN DESIGN SPECIFICATION. THE INSTRUMENT WAS DISASSEMBLED TO EXAMINE THE INTERNAL COMPONENTS AND NO DEFORMATIONS COULD BE IDENTIFIED. IT WAS CONCLUDED THAT THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO DESIGN SPECIFICATIONS. THE EXPERIENCE THE SURGEON REPORTED COULD NOT BE REPEATED. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE IT FUNCTIONS PROPERLY. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE CO'S PRODUCTS.
IT WAS REPORTED THE EZ45 WAS USED DURING A THORACOSCOPY. IT WAS REPORTED THE SURGEON INSERTED THE EZ45 INTO A MINITHORACOTOMY. THE JAWS WERE OPENED AND THE RELOAD FELL INTO PATIENT. THE RELOAD WAS RETRIEVED AND INSTRUMENT DISCARDED. ANOTHER EZ45 WAS OPENED AND WORKED FINE. THERE WAS NO CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAFETY LOCK | ENDOSCOPIC LINEAR CUTTER | GAG | ETHICON ENDO-SURGERY, INC. | NA | J45L8E |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |