FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1100024 · Received August 5, 2008

Report

Report Number
2531779-2008-00456
Event Type
Malfunction
Date Received
August 5, 2008
Date of Event
July 8, 2008
Report Date
July 8, 2008
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED AN INTERMITTENT LOSS OF CONTACT BETWEEN THE BATTERY CAP AND THE PUMP CASE.

Description of Event or Problem · 1

THE PT REPORTED THAT THE PUMP WAS RESETTING ITSELF WITHOUT INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. IR1250

Patients

Seq Age Sex Outcome Treatment
1