Report Number
2531779-2008-00455
Event Type
Malfunction
Date Received
August 5, 2008
Date of Event
July 8, 2008
Report Date
July 8, 2008
Manufacturer
ANIMAS CORP.
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED AN INTERMITTENT LOSS OF CONTACT BETWEEN THE BATTERY CAP AND THE PUMP CASE.

Description of Event or Problem · 1

THE PT REPORTED THAT THE PUMP WAS RESETTING ITSELF WITHOUT INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. IR1250

Patients

Seq Age Sex Outcome Treatment
1