FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT II 10 ML SYRINGE

MDR report key: 10997376 · Received December 13, 2020

Report

Report Number
3002682307-2020-00401
Event Type
Malfunction
Date Received
December 13, 2020
Date of Event
November 16, 2020
Report Date
March 18, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2003190. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR RETURN TWENTY RETAINED SAMPLES OF THE SAME LOT NUMBER WERE OBTAINED FROM THE MANUFACTURING FACILITY FOR FURTHER EVALUATION. THE RETAINED SAMPLES WERE EXAMINED AND NO DEFECTS WERE OBSERVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD DISCARDIT¿ II 10 ML SYRINGE EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER INFORMS THAT THE PLUNGER OF THE 10 ML DISCARDIT SYRINGE IS LEAKING WHEN INJECTING THE MEDICINE AND SPILLS OVER THE HANDS OF THE NURSE. THE MEDICINE DOES NOT STAY STERILE IN THE SYRINGE.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE LOT #: AN INVALID LOT # OF 2003190 WAS PROVIDED BY THE INITIAL REPORTER. DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD DISCARDIT¿ II 10 ML SYRINGE EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER INFORMS THAT THE PLUNGER OF THE 10 ML DISCARDIT SYRINGE IS LEAKING WHEN INJECTING THE MEDICINE AND SPILLS OVER THE HANDS OF THE NURSE. THE MEDICINE DOES NOT STAY STERILE IN THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1462382 BD DISCARDIT II 10 ML SYRINGE SYRINGE FMF BECTON DICKINSON, S.A. UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1