FDA Adverse Event Malfunction Summary report: N

ORTHO. MALLET 2 LB/ 1 LB

MDR report key: 10993326 · Received December 11, 2020

Report

Report Number
3011137372-2020-00287
Event Type
Malfunction
Date Received
December 11, 2020
Date of Event
December 3, 2020
Report Date
December 3, 2020
Manufacturer
TELEFLEX MEDICAL
Product Code
HXL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

QN#(B)(4). DEVICE HISTORY RECORD (DHR) WAS REVIEWED FOR NOTIFICATIONS PERTAINING TO INCOMING INSPECTION, STOCK CHECKS, AND PRODUCT RETURNS. NO CONCERNS WERE NOTED WITH REFERENCE TO KM46666. (1) SAMPLE OF KM46666 LOT A9 WAS RECEIVED FOR EVALUATION. THE MALLET HEAD HAS BROKEN AWAY FROM THE HANDLE. THE FRACTURE PLANE SHOWS CRACKS PARALLEL TO THE IMPACT PLANE WHICH IS ABNORMAL. THE MALLET FACE EXHIBITS DEFORMATION AND THE EDGES OF THE MALLET FACE FROM EXCESSIVE IMPACT. DAMAGE IS CONSISTENT WITH OVER STRESSING OF THE MALLET HEAD. THE MALLET FACE IS DEFORMED INDICATING THAT EXCESSIVE FORCE HAS BEEN APPLIED OFF CENTER FROM THE MALLET FACE. (1) SAMPLE OF KM46666 LOT A9 WAS RECEIVED FOR EVALUATION. THE MALLET HEAD HAS BROKEN AWAY FROM THE HANDLE. THE FRACTURE PLANE SHOWS CRACKS PARALLEL TO THE IMPACT PLANE WHICH IS ABNORMAL. THE MALLET FACE EXHIBITS DEFORMATION AND THE EDGES OF THE MALLET FACE FROM EXCESSIVE IMPACT. A DIMENSIONAL INSPECTION WAS NOT REQUIRED AS PART OF THIS INVESTIGATION. A FUNCTIONAL INSPECTION WAS NOT REQUIRED AS PART OF THIS INVESTIGATION. DEVICE HISTORY RECORD (DHR) WAS REVIEWED FOR NOTIFICATIONS PERTAINING TO INCOMING INSPECTION, STOCK CHECKS, AND PRODUCT RETURNS. NO CONCERNS WERE NOTED WITH REFERENCE TO KM46666 DAMAGE IS CONSISTENT WITH OVER STRESSING OF THE MALLET HEAD. THE MALLET FACE IS DEFORMED INDICATING THAT EXCESSIVE FORCE HAS BEEN APPLIED OFF CENTER FROM THE MALLET FACE.

Description of Event or Problem · 0

REPORTED ISSUE: I NEED TO SEND IN A MALLET THAT BROKE DURING A CASE. THE HEAD SHEARED OFF AT HANDLE AGAIN.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATED EVENTS.

Description of Event or Problem · 1

REPORTED ISSUE: I NEED TO SEND IN A MALLET THAT BROKE DURING A CASE. THE HEAD SHEARED OFF AT HANDLE AGAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1454929 ORTHO. MALLET 2 LB/ 1 LB MALLET HXL TELEFLEX MEDICAL N/A

Patients

Seq Age Sex Outcome Treatment
1