FDA Adverse Event Death Summary report: N

PALL SMALL VOLUME HEAT & MOISTURE EXCHANGER

MDR report key: 109914 · Received August 4, 1997

Report

Report Number
9680602-1997-00001
Event Type
Death
Date Received
August 4, 1997
Date of Event
July 10, 1997
Report Date
July 11, 1997
Manufacturer
PALL NEWQUAY, LTD.
Product Code
BYD
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION/ANALYSIS THE RETURNED DEVICE WAS FOUND TO BE REPRESENTATIVE OF CURRENT MFG PRODUCT AND EXHIBITED NORMAL FLOW/PRESSURE CHARACTERISTICS WHEN TESTED. THIS IS IN AGREEMENT WITH THE OBSERVATION OF THE ICU CHIEF, WHO TESTED THE DEVICE WITH A RESERVOIR BAG 10-12 HOURS AFTER THE INVESTIGATION. IN THE FIRM'S OPINION, THE RETURNED DEVICE HAD NOT MALFUNCTIONED IN ANY WAY AND DID NOT CONTRIBUTE TO THE PT'S DEATH. FURTHERMORE, THE MEDICAL EXAMINER, HAS VERBALLY INFORMED REPRESENTATIVES OF THE FIRM THAT THE PT HAD MYOCARDIAL INFARCTION OF THE WALL OF THE LEFT VENTRICLE AND THAT THE DEVICE DID NOT CONTRIBUTE TO THE PT'S DEATH. ADDITIONL INFO: CO HAS WRETTEN TO THE PATHOLOGIST FOR ADDITIONAL INFO AND TO CONFIRM CO'S STATEMENT, BUT HAVE NOT RECEIVED A RESPONSE. UNLESS SUBSTANTIALLY SIGNIFICANT INFO BECOMES AVAILABLE, THIS CONSTITUTES A FINAL REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PT WITH COPD WAS ADMITTED TO THE HOSPITAL'S EMERGENCY DEPT DUE TO HYPONATRAEMIA AND LOSS OF CONSCIOUSNESS. THE PT WAS CONNECTED TO A IMV-BIRD VENTILATOR. A PALL BB25 DEVICE WAS PRESENT AT THE PT END. VENTOLIN AND ATROVENT WERE NEBULIZED TO THE PT. TEN (10) MINUTES AFTER NEBULIZATION IT WAS NOTICED THAT THE PT WAS CYANOTIC. ATTENTION WAS DRAWN TO THE FACT THAT THE PT'S CHEST WAS NOT MOVING IN RESPONSE TO THE VENTILATOR. ATTEMPTS TO RESUSCIATE THE PT WERE UNSUCCESSFUL. THE PT EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PALL SMALL VOLUME HEAT & MOISTURE EXCHANGER SMALL VOLUME HEAT & MOISTURE EXCHANGER BYD PALL NEWQUAY, LTD. BB25 701501

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Death