FDA Adverse Event Malfunction Summary report: N

BD SYRINGE LUER-LOK TIP

MDR report key: 10988571 · Received December 10, 2020

Report

Report Number
1213809-2020-00884
Event Type
Malfunction
Date Received
December 10, 2020
Date of Event
November 12, 2020
Report Date
December 29, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 12/14/2020. H.6. INVESTIGATION: ONE LOOSE 10ML SYRINGE WAS RECEIVED AND VISUALLY EVALUATED. THE BARREL WAS OBSERVED TO HAVE ONE CRACK ALONG THE BARREL¿S LENGTH BETWEEN 6ML AND BD LOGO MARKINGS. THE CRACK WAS UNDER THE NUMERICAL SCALE MARKINGS. POTENTIAL ROOT CAUSE FOR THE CRACKED BARREL DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 DIFFERENT BD SYRINGES WITH THE LUER-LOK¿ TIP HAD CRACKS IN THEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM DUTCH TO ENGLISH: "THERE IS A CRACK IN 2 SYRINGES."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9081846, MEDICAL DEVICE EXPIRATION DATE: 2024-03-31, DEVICE MANUFACTURE DATE: 2019-03-22. MEDICAL DEVICE LOT #: 9108747, MEDICAL DEVICE EXPIRATION DATE: 2024-03-31, DEVICE MANUFACTURE DATE: 2019-04-18. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT 2 DIFFERENT BD SYRINGES WITH THE LUER-LOK" TIP HAD CRACKS IN THEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "THERE IS A CRACK IN 2 SYRINGES."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1447004 BD SYRINGE LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS SEE SECTION H.10.

Patients

Seq Age Sex Outcome Treatment
1