MOBILE INFANT WARMER
Report
- Report Number
- 9611451-2008-00476
- Event Type
- Malfunction
- Date Received
- August 5, 2008
- Report Date
- July 8, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- FMT
- PMA / PMN Number
- K970432
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
AN INITIAL INVESTIGATION WAS CARRIED OUT BASED ON THE EVENT DESCRIPTION AND RESULTS OF PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS, AS THE COMPLAINT DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION. RESULTS: CRACKING OF THE INFANT WARMER HEAD IS USUALLY DUE TO THE USE OF INCOMPATIBLE CLEANING SOLUTIONS AND DOING THE CLEANING BEFORE THE WARMER HEAD HAS BEEN COOLED. IT WAS REPORTED THAT THE HOSPITAL WAS USING A MIXTURE OF PRECISE QTB, ALKYL DIMETHYL BENZYL, AMMONIUM CHLORIDE AND ETHYL BENZYL BENZOYL IN CLEANING THE INFANT WARMER HEAD. THE OPERATING MANUAL STATES THAT CLEANING OF ALL PARTS OF THE INFANT WARMER AND ACCESSORIES SHOULD BE DONE AT NORMAL ROOM TEMPERATURE, AROUND 23 DEGRESS C. THE DEVICE IS MADE UP OF POLYCARBONATE PLASTICS AND CLEANING SOLUTIONS CONTAINING AROMATIC HYDROCARBONS, AMMONIA, AMINES OR AQUEOUS SOLUTIONS ARE NOT RECOMMENDED AS THESE MAY CAUSE CHEMICAL REACTIONS TO THE PLASTIC MATERIAL. ALL SURFACES OF THE WARMER HEAD MAY BE CLEANED WITH DETERGENT BASED SOLUTION WITH MAXIMUM CONCENTRATION OF 2% IN WATER. CONCLUSION: CRACKING OF THE INFANT WARMER HEAD, DUE TO THE USE OF NON-COMPATIBLE CLEANING MATERIALS, IS A KNOWN PROBLEM. WE HAVE AN OPEN CAPA TO EXAMINE FURTHER COMPATIBLE CLEANING SOLUTION.
A HOSPITAL IN CARBONDALE REPORTED THAT THEIR MOBILE INFANT WARMER HEAD HAD CRACKED, AS THEY WERE USING THE WRONG CLEANING PRODUCT. THEY WERE USING A MIXTURE OF PRECISE QTB, ALKYL DIMETHYL BENZYL, AMMONIUM CHLORIDE AND ETHYL BENZYL BENZOYL IN CLEANING THE INFANT WARMER HEAD. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MOBILE INFANT WARMER | FMT | FISHER & PAYKEL HEALTHCARE, LTD. | IW910JEU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |