BD RAPID DETECTION OF SARS-COV-2 VERITOR¿
Report
- Report Number
- 1119779-2020-01197
- Event Type
- Malfunction
- Date Received
- December 10, 2020
- Date of Event
- November 22, 2020
- Report Date
- November 26, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- QKP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION: THIS STATEMENT SUMMARIZES THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE POSITIVE RESULTS WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR (MATERIAL # 256082), BATCH 0233975. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE POSITIVE COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. INVESTIGATIONS WERE PERFORMED ON THE BATCH NUMBER 0233975 AND NO RELEVANT ISSUE WAS FOUND. RETENTION SAMPLE TESTING WAS PERFORMED ON BATCH NUMBER 0233975; NO RELEVANT ISSUE WAS FOUND. THE COMPLAINT WAS UNABLE TO BE CONFIRMED VIA THE RETAINED SAMPLES. A TREND WAS IDENTIFIED FOR FALSE POSITIVE RESULTS. BASED ON THE TREND, CAPA#1878253 WAS INITIATED. BD QUALITY WILL CONTINUE TO MONITOR FOR TRENDS. THERE WERE NO CORRECTIVE ACTIONS TAKEN AT THIS TIME.
IT WAS REPORTED WHILE TESTING FOR SARS-COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING PCR TEST METHOD AND THE RESULT WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT. CUSTOMER HAS STATED NO ADDITIONAL DETAILS WILL BE PROVIDED. EUA #(B)(4).
EUA # (B)(4). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED WHILE TESTING FOR SARS-COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING PCR TEST METHOD AND THE RESULT WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT. CUSTOMER HAS STATED NO ADDITIONAL DETAILS WILL BE PROVIDED. EUA # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1450321 | BD RAPID DETECTION OF SARS-COV-2 VERITOR¿ | CORONAVIRUS ANTIGEN DETECTION SYSTEM | QKP | BECTON, DICKINSON & CO. (SPARKS) | 0233975 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |