HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2020-00677
- Event Type
- Malfunction
- Date Received
- December 10, 2020
- Date of Event
- November 11, 2020
- Report Date
- December 10, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). LIVANOVA FIELD SERVICE REPRESENTATIVE DISPATCHED TO THE CUSTOMER'S FACILITY WAS ABLE TO CONFIRM THE ISSUE. TO SOLVE THE ISSUE THE STIRRER MOTOR AND THE DISTRIBUTION BOARD WERE REPLACED. SUBSEQUENT FUNCTIONAL VERIFICATION TESTING WAS COMPLETED WITHOUT FURTHER ISSUES AND THE UNIT WAS RETURNED TO SERVICE.
LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DISPLAYED AN ERROR MESSAGE ASSOCIATED TO THE STIRRER MOTOR DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1453600 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |