FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10982757 · Received December 10, 2020

Report

Report Number
9611109-2020-00677
Event Type
Malfunction
Date Received
December 10, 2020
Date of Event
November 11, 2020
Report Date
December 10, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). LIVANOVA FIELD SERVICE REPRESENTATIVE DISPATCHED TO THE CUSTOMER'S FACILITY WAS ABLE TO CONFIRM THE ISSUE. TO SOLVE THE ISSUE THE STIRRER MOTOR AND THE DISTRIBUTION BOARD WERE REPLACED. SUBSEQUENT FUNCTIONAL VERIFICATION TESTING WAS COMPLETED WITHOUT FURTHER ISSUES AND THE UNIT WAS RETURNED TO SERVICE.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DISPLAYED AN ERROR MESSAGE ASSOCIATED TO THE STIRRER MOTOR DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.