FDA Adverse Event Malfunction Summary report: N

630G INSULIN PUMP MMT-1715K 630G BLACK MG

MDR report key: 10981598 · Received December 10, 2020

Report

Report Number
2032227-2020-215376
Event Type
Malfunction
Date Received
December 10, 2020
Date of Event
December 4, 2020
Report Date
October 18, 2021
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
000000763000316631
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE RECEIVED WITH NO RESPONSE MENU KEYPAD BUTTON. CUT AND OPEN THE DEVICE FOUND MOISTURE/DAMAGE ON THE ELECTRICAL BOARD 1, INTERNAL BATTERY CONNECTOR DURING VISUAL INSPECTION. PERFORM BRUSH CLEANING WITH THE ISOPROPYL ALCOHOL AND CLEANING UP ALL CORROSION ON ELECTRICAL BOARD 1 AND POWERED THE INSULIN PUMP ON USING THE BATTERY SIMULATOR NORMALLY AND DEVICE STILL NO RESPONSE. DEVICE WAS TESTED WITH TEST KEYPAD TRACES AND ALL KEYPAD FUNCTION PROPERLY. IN CONCLUSION, NO RESPONSE MENU KEYPAD BUTTON DUE TO MOISTURE DAMAGE ELECTRICAL BOARD 1. DEVICE HAD SCRATCHED CASE, CRACKED KEYPAD OVERLAY. THE DEVICE P-CAP / TEST RESERVOIR LOCKS IN PLACE PROPERLY AND NO ANOMALY NOTED. DEVICE HAD CRACKED KEYPAD OVERLAY, SCRATCHED CASE. THE DEVICE P-CAP / TEST RESERVOIR LOCKS IN PLACE PROPERLY AND NO ANOMALY NOTED. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BUTTONS WAS UNRESPONSIVE AND INSULIN PUMP WAS LEAKING INSULIN AND SCREEN WAS NOT LAST AS LONG WHEN TURNED ON. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP AND RESERVOIR WILL NOT BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1453825 630G INSULIN PUMP MMT-1715K 630G BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1715K HG44J0L 000000763000316631

Patients

Seq Age Sex Outcome Treatment
1 63 YR