FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE INSULIN SYRINGE

MDR report key: 10976225 · Received December 9, 2020

Report

Report Number
1920898-2020-01687
Event Type
Malfunction
Date Received
December 9, 2020
Date of Event
November 13, 2020
Report Date
March 9, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTOS PROVIDED. 3 PHOTOS OF (3) LOOSE 0.3ML BD INSULIN SYRINGES WERE PROVIDED. THE CUSTOMER REPORTED WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL. THE PHOTOS WERE REVIEWED, AND IT WAS OBSERVED THAT 2 OUT OF THE 3 SYRINGES EXHIBITED A NEEDLE HUB/SHIELD ASSEMBLY SEPARATED FROM THE BARREL; NO DAMAGE TO THE BARREL TIPS WAS OBSERVED. THE OTHER 1 SYRINGE EXHIBITED A HUB ASSEMBLY NOT FULLY SEATED ON THE BARREL TIP. DUE TO BATCH NUMBER BEING UNKNOWN, NO DHR REVIEW CAN BE COMPLETED. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CORRECTIVE/PREVENTATIVE ACTION (CAPA#: 1630423) HAS BEEN INITIATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE HUB SEPARATED FROM THE BD ULTRA-FINE¿ INSULIN SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: "WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL."

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NEEDLE HUB SEPARATED FROM THE BD ULTRA-FINE¿ INSULIN SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL."