FDA Adverse Event Other Summary report: N

BAXTER

MDR report key: 109740 · Received July 28, 1997

Report

Report Number
109740
Event Type
Other
Date Received
July 28, 1997
Date of Event
June 20, 1997
Report Date
June 24, 1997
Manufacturer
BAXTER HEALTHCARE CORP.
Product Code
FRN
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
ME, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

NTG DRIP INSTALLED. STAFF FAILED TO REMOVE CLAMP FROM LINE ON SIDE 1 OF DOUBLE PUMP. AND NO NTG DELIVERED. PUMP COUNTED DOWN TO EMPTY IN SPITE OF THIS, BUT FAILED TO BEEP OCCLUSION. IV BOTTLE REMAINED FULL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BAXTER DUAL CHANNEL VOLUMETRIC INFUSION PUMP FRN BAXTER HEALTHCARE CORP. FLO-GARD 6300 *

Patients

Seq Age Sex Outcome Treatment
1 69 YR Other