TITAN
Report
- Report Number
- 2125050-2020-01229
- Event Type
- Malfunction
- Date Received
- December 9, 2020
- Report Date
- March 1, 2021
- Manufacturer
- COLOPLAST A/S
- Product Code
- FHW
- PMA / PMN Number
- P000006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A TITAN TOUCH PUMP WAS RECEIVED FOR EVALUATION. EXAMINATION AND TESTING OF THE RETURNED COMPONENTS REVEALED NO FUNCTIONAL ABNORMALITIES. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. A REVIEW OF THE COMPLAINT HISTORY DATABASE, NON-CONFORMANCES AND CAPAS REVEALED NO TRENDS FOR THIS LOT.
WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. ONCE OUR EVALUATION IS COMPLETE, A FOLLOW-UP REPORT WILL BE SUBMITTED. COLOPLAST HAS NOT BEEN PROVIDED ANY CORROBORATING EVIDENCE TO VERIFY THE INFORMATION CONTAINED IN THIS REPORT.
ACCORDING TO THE AVAILABLE INFORMATION, THOUGH NOT VERIFIED, REPORTED COMPLAINT: "THE PUMP IS HARD AND IT IS SLOW TO INFLATE THE PROTHESIS AND TO DEFLATE IS VIRTUALLY IMPOSSIBLE. HE HAS TO USE EXTREME FORCE AND TRY 4 OR 5 TIMES WHICH IS VERY PAINFUL TO DO. IT TAKES AROUND 5 ¿ 10 MINS TO GET IT STARTED TO DEFLATE. WHEN THE PATIENT ARRIVED AT [PHYSICIAN] IT WAS STILL PARTIALLY INFLATED AS HE COULD NOT GET IT DOWN COMPLETELY. [PHYSICIAN] HAS TRIED THE TROUBLESHOOTING TECHNIQUES WITH NO SUCCESS. THE DEVICE WAS REVISED/REPLACED. WILL BE RETURNED. ADDITIONAL INFORMATION RECEIVED 11/15/2020: ONLY THE PUMP WAS REPLACED. ADDITIONAL INFORMATION RECEIVED 11/118/2020: EXPLANTED DEVICE BEING RETURNED. THE LOT # WAS REVIEWED FOR COMPLAINT TREND, NONCONFORMING REPORT AND CAPA. DEVICES MET SPECIFICATION PRIOR TO RELEASE AND NO TRENDS WERE NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1442111 | TITAN | INFLATABLE PENILE PROTHESIS | FHW | COLOPLAST A/S | ES29181022 | 6401130 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |