BEARING +3 MM THICKNESS 36 MM DIAMETER
Report
- Report Number
- 0001822565-2020-03985
- Event Type
- Injury
- Date Received
- December 8, 2020
- Date of Event
- November 17, 2020
- Report Date
- March 24, 2021
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- PHX
- PMA / PMN Number
- K181611
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED. VISUAL EXAMINATION OF THE RETURNED PRODUCT IDENTIFIED THAT THE BEARING +3 MM THICKNESS 36 MM DIAMETER LOT 64343811 EXHIBITED HEAVY WEAR ON ALL SURFACES AND ACCURATE DIMENSIONAL ANALYSIS WAS NOT POSSIBLE. REVIEW OF THE DEVICE HISTORY RECORD(S) IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
(B)(4). PROCODE: PHX. CONCOMITANT MEDICAL PRODUCTS: MINI HUMERAL TRAY STANDARD THICKNESS +0 MM TAPER OFFSET 40 MM DIAMETER, CAT#: 110031399, LOT#: 64666139. COMP RVRS SHLDR GLNSP STD 36MM, CAT#: 115310, LOT#: 886340. 25MM VERSA-DIAL TAPER ADAPTOR, CAT#:118000-00, LOT#: 932950. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001822565 - 2020 - 03984.
IT WAS REPORTED THE PATIENT UNDERWENT A SHOULDER PROCEDURE. APPROXIMATELY 6 MONTHS LATER, THE PATIENT WAS REVISED DUE TO PAIN, CLICKING AND DISASSOCIATION. DURING THE REVISION, THE SURGEON NOTICED THE POLYETHYLENE HAD DISASSOCIATED FROM THE HUMERAL TRAY. THE GLENOSPHERE HAD ALSO COME IN CONTACT WITH THE HUMERAL TRAY AND HAD BEEN DAMAGED. THE DOCTOR REPLACED THE GLENOSPHERE, TAPER ADAPTOR, HUMERAL POLY, AND HUMERAL TRAY. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1430497 | BEARING +3 MM THICKNESS 36 MM DIAMETER | PROSTHESIS, SHOULDER | PHX | ZIMMER BIOMET, INC. | NI | 64343811 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |