FDA Adverse Event Malfunction Summary report: N

8110 ALARIS SYRINGE PUMP

MDR report key: 10964053 · Received December 8, 2020

Report

Report Number
2016493-2020-60013
Event Type
Malfunction
Date Received
December 8, 2020
Report Date
March 11, 2019
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403424267
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACK WISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

BROKEN/DAMAGED DAMAGED/CRACKED BARREL CLAMP ASSY ISOLATION RIB DAMAGE DAMAGED/CRACKED HANDLE TUBE DRIVE BENT DAMAGED/CRACKED FLANGE GRIPPER WIPER SEAL DAMAGED DAMAGED/CRACKED REAR & FRONT CASE 03/11/2019 09:53:13 RFC_REPAIRS (RFC_REPAIRS) PO FOR $579. 03/28/2019 21:32:46 DIANE H NGUYEN (DHNGUYEN) REPLACED FRONT CASE, REAR CASE, HANDLE LEFT, FLANGE GRIPPER, TUBE DRIVEARM, SEAL WIPER, IUI RIGHT CONNETOR, AND BARREL CLAMP. PERFORMED PM, AND CALIBRATION. 03/28/2019 21:48:05 DIANE H NGUYEN (DHNGUYEN) RECEIVED THIS DEVICE WITH SOFTWARE V9.33.0.50. 03/29/2019 07:58:58 ANNETTE A MENDEZ (AMENDEZ) 9632001960920413730700480982914357

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1432680 8110 ALARIS SYRINGE PUMP PUMP, INFUSION FRN CAREFUSION SD 8110 10885403424267

Patients

Seq Age Sex Outcome Treatment
1