8110 ALARIS SYRINGE PUMP
Report
- Report Number
- 2016493-2020-60013
- Event Type
- Malfunction
- Date Received
- December 8, 2020
- Report Date
- March 11, 2019
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403424267
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACK WISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
BROKEN/DAMAGED DAMAGED/CRACKED BARREL CLAMP ASSY ISOLATION RIB DAMAGE DAMAGED/CRACKED HANDLE TUBE DRIVE BENT DAMAGED/CRACKED FLANGE GRIPPER WIPER SEAL DAMAGED DAMAGED/CRACKED REAR & FRONT CASE 03/11/2019 09:53:13 RFC_REPAIRS (RFC_REPAIRS) PO FOR $579. 03/28/2019 21:32:46 DIANE H NGUYEN (DHNGUYEN) REPLACED FRONT CASE, REAR CASE, HANDLE LEFT, FLANGE GRIPPER, TUBE DRIVEARM, SEAL WIPER, IUI RIGHT CONNETOR, AND BARREL CLAMP. PERFORMED PM, AND CALIBRATION. 03/28/2019 21:48:05 DIANE H NGUYEN (DHNGUYEN) RECEIVED THIS DEVICE WITH SOFTWARE V9.33.0.50. 03/29/2019 07:58:58 ANNETTE A MENDEZ (AMENDEZ) 9632001960920413730700480982914357
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1432680 | 8110 ALARIS SYRINGE PUMP | PUMP, INFUSION | FRN | CAREFUSION SD | 8110 | 10885403424267 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |