BD INSYTE AUTOGUARD BC SHIELDED IV CATHETER
Report
- Report Number
- 1710034-2020-00784
- Event Type
- Malfunction
- Date Received
- December 7, 2020
- Date of Event
- October 4, 2020
- Report Date
- January 20, 2021
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- UDI-DI
- 00382903825233
- PMA / PMN Number
- K110443
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 12/3/2020. H.6. INVESTIGATION: BD RECEIVED 2,242 UNOPENED UNITS. PER BD POLICY, FORTY-FIVE UNITS WERE SELECTED AT RANDOM FOR TESTING. THE UNITS WERE TESTED FOR LEAKAGE BEYOND THE SEPTUM BUT NO LEAKAGE WAS OBSERVED. BD WAS UNABLE TO CONFIRM THE REPORTED DEFECT. IF THE SEPTUM IS ENGAGED WHEN THE CONNECTION IS MADE (OR BEFORE) BLOOD WILL FLOW THROUGH THE OPENED SEPTUM, AS NOTED IN THE IFU. THE IFU ALSO STATES THAT THE BLOOD CONTROL FEATURE IS NOT INTENDED FOR LONG TERM USE AND A SECURE CONNECTION SHOULD BE MADE WITHIN 10 SECONDS. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT THE BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER EXPERIENCED BLOOD LEAKAGE THROUGH THE BC VALVE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO : 382523, BATCH NO: 0252150. BLOOD CONTROL SEPTUM NOT EFFECTIVELY CONTAINING BLOOD. ONCE IV STARTED, BLOOD SPEWS OUT OF HUB IMMEDIATELY RATHER THAN STAYING CONTAINED BEHIND THE SEPTUM.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT THE BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER EXPERIENCED BLOOD LEAKAGE THROUGH THE BC VALVE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO :382523 BATCH NO: 0252150. BLOOD CONTROL SEPTUM NOT EFFECTIVELY CONTAINING BLOOD. ONCE IV STARTED, BLOOD SPEWS OUT OF HUB IMMEDIATELY RATHER THAN STAYING CONTAINED BEHIND THE SEPTUM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1428684 | BD INSYTE AUTOGUARD BC SHIELDED IV CATHETER | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 0252150 | 00382903825233 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |