FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP

MDR report key: 10963687 · Received December 7, 2020

Report

Report Number
2243072-2020-02015
Event Type
Malfunction
Date Received
December 7, 2020
Date of Event
November 12, 2020
Report Date
March 9, 2021
Manufacturer
BECTON DICKINSON
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTOS PROVIDED. 2 PHOTOS OF (4) LOOSE 0.3ML BD INSULIN SYRINGES WERE PROVIDED. THE CUSTOMER REPORTED WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIELD WAS SEPARATED FROM THE BARREL. THE PHOTOS WERE REVIEWED, AND IT WAS OBSERVED THAT 2 OUT OF THE 4 SYRINGES EXHIBITED HUB SEPARATION. HUB SEPARATION WAS NOT OBSERVED ON THE OTHER 2 SYRINGES PICTURED. DUE TO BATCH NUMBER BEING UNKNOWN, NO DHR REVIEW CAN BE COMPLETED. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA#1630423 WAS INITIATED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP SEPARATED FROM THE HUB DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL."

Additional Manufacturer Narrative · 1

UNKNOWN MANUFACTURER: (B)(4). MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP SEPARATED FROM THE HUB DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1427937 SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP PISTON SYRINGE FMF BECTON DICKINSON UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1