FDA Adverse Event Malfunction Summary report: N

BD 1 ML SYRINGE LUER SLIP TIP CONVENIENCE PAK

MDR report key: 10962197 · Received December 7, 2020

Report

Report Number
9610847-2020-00397
Event Type
Malfunction
Date Received
December 7, 2020
Date of Event
October 29, 2020
Report Date
February 18, 2021
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMF
UDI-DI
00382903097012
PMA / PMN Number
K980580
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR PROVIDED FINAL LOT NUMBER 0092223 AND ALL APPLICABLE SUB-ASSEMBLY LOT NUMBERS. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED INCIDENT. TO AID IN THE INVESTIGATION OF THIS ISSUE, BOTH PICTURE AND PHYSICAL SAMPLES WERE RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE SAMPLES, THE BARREL CAVITY OF THE SYRINGE WAS OBSERVED DAMAGED. THE DAMAGES APPEARED TO BE ELONGATED AND INDENTED. IT HAS BEEN DETERMINED THAT THE DAMAGED BARRELS MOST LIKELY RESULTED DURING THE BD ASSEMBLY PROCESS; HOWEVER, AND EXACT POINT OF ERROR HAS NOT BEEN IDENTIFIED. SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 25 BD 1 ML SYRINGE LUER SLIP TIP CONVENIENCE PAKS EXPERIENCED MOLDING DEFECTS -SHORT SHOTS ON BARREL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 309701 BATCH NO: 0092223 I'M REACHING OUT IN REGARDS TO WHAT APPEARS TO BE A MOLDING ISSUE WITH OUR 1ML SLIP TIP SYRINGE'S. THIS PARTICULAR ISSUE IS FOUND IN THE 25 COUNT TRAYS THAT WE ORDER. THE DEFECT APPEARS TO BE A SMALL INDENTATION AND IS ABOUT 0.5 CM IN LENGTH, IT IS THE SAME SIZE AND SHAPE ON EVERY AFFECTED SYRINGE THAT WE ENCOUNTER LEADING US TO BELIEVE IT COULD BE AN ERROR WITH THE MOLDING.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT 25 BD 1 ML SYRINGE LUER SLIP TIP CONVENIENCE PAKS EXPERIENCED MOLDING DEFECTS SHORT SHOTS ON BARREL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 309701, BATCH NO: 0092223. I'M REACHING OUT IN REGARDS TO WHAT APPEARS TO BE A MOLDING ISSUE WITH OUR 1ML SLIP TIP SYRINGE'S. THIS PARTICULAR ISSUE IS FOUND IN THE 25 COUNT TRAYS THAT WE ORDER. THE DEFECT APPEARS TO BE A SMALL INDENTATION AND IS ABOUT 0.5 CM IN LENGTH, IT IS THE SAME SIZE AND SHAPE ON EVERY AFFECTED SYRINGE THAT WE ENCOUNTER LEADING US TO BELIEVE IT COULD BE AN ERROR WITH THE MOLDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1426863 BD 1 ML SYRINGE LUER SLIP TIP CONVENIENCE PAK PISTON SYRINGE FMF BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 0092223 00382903097012

Patients

Seq Age Sex Outcome Treatment
1