FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 1096211 · Received July 31, 2008

Report

Report Number
1644487-2008-01747
Event Type
Malfunction
Date Received
July 31, 2008
Date of Event
July 2, 2008
Report Date
July 2, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

REPORTER INDICATED THAT AT A PT'S INITIAL DOSING APPOINTMENT, THE VNS GENERATOR WAS FOUND TO BE AT 1MA, 20HZ, 30 SECONDS ON TIME, AND 60 MINUTES OFF TIME. THE PHYSICIAN REPORTED THAT THERE WAS AN INTERRUPTED SYSTEM DIAGNOSTICS TEST AT THE OFFICE VISIT; HOWEVER, IT IS UNCLEAR IF THE UNINTENDED SETTINGS OCCURRED DUE TO THAT TEST OR BEFORE THE OFFICE VISIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. 250

Patients

Seq Age Sex Outcome Treatment
1 20 YR