FDA Adverse Event Malfunction Summary report: N

BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY

MDR report key: 10959840 · Received December 7, 2020

Report

Report Number
1119779-2020-01145
Event Type
Malfunction
Date Received
December 7, 2020
Date of Event
September 9, 2020
Report Date
September 21, 2021
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
QKP
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: THIS IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE POSITIVE RESULTS WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR (MATERIAL # 256082), BATCH NUMBER UNKNOWN. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE POSITIVE COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. AN INVESTIGATION COULD NOT BE PERFORMED AS NO BATCH NUMBER OR AN INCORRECT BATCH NUMBER WAS PROVIDED. THE COMPLAINT WAS UNABLE TO BE CONFIRMED. HOWEVER, THERE IS A TREND AGAINST FALSE POSITIVE RESULTS. BD HAS INITIATED CAPA (CORRECTIVE AND PREVENTIVE ACTION) # 1878253 TO FURTHER INVESTIGATE. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. A REPEAT TEST WAS PERFORMED USING PCR AND THE RESULT WAS NEGATIVE. THE CUSTOMER STATED THEY ARE TESTING ASYMPTOMATIC PATIENTS. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THE CUSTOMER STATED RESULTS WERE REPORTED OUT, BUT THERE WAS NO PATIENT IMPACT. EUA#: (B)(4).

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4). MEDICAL DEVICE LOT #: 0220417 WAS REPORTED, HOWEVER, THIS IS NOT A LOT# MANUFACTURED FOR THIS PRODUCT #. FDA NOTIFIED?: THE INITIAL REPORTER ALSO NOTIFIED THE FDA ON (DATE) VIA MEDWATCH # MW5097003.

Description of Event or Problem · 1

IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. A REPEAT TEST WAS PERFORMED USING PCR AND THE RESULT WAS NEGATIVE. THE CUSTOMER STATED THEY ARE TESTING ASYMPTOMATIC PATIENTS. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THE CUSTOMER STATED RESULTS WERE REPORTED OUT, BUT THERE WAS NO PATIENT IMPACT. EUA#: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1426086 BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY CORONAVIRUS ANTIGEN DETECTION SYSTEM QKP BECTON, DICKINSON & CO. (SPARKS) 0220417

Patients

Seq Age Sex Outcome Treatment
1