FDA Adverse Event Injury Summary report: N

LINER: MPACT DM DOUBLE MOBILITY HC LINER 22.2/DMC

MDR report key: 10959367 · Received December 7, 2020

Report

Report Number
3005180920-2020-00878
Event Type
Injury
Date Received
December 7, 2020
Date of Event
November 10, 2020
Report Date
December 7, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
MEH
UDI-DI
07630030806834
PMA / PMN Number
K092265
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 20 NOVEMBER 2020: LOT 184721: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26-JULY-2018. EXPIRATION DATE: 2023-07-15. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT CLINICAL EVALAUTION: HIP DISLOCATION 3 MONTHS AFTER IMPLANTATION OF A DOUBLE MOBILITY IMPLANT IN A CHALLENGING CASE DUE TO PATIENT CLINICAL HISTORY OF A SURGICALLY TREATED INTERTROCHANTERIC FRACTURE. DISLOCATION PROBABLY OCCURRED DURING REHABILITATION. HIP DISLOCATION MAY BE DUE TO COMPONENT POSITIONING OR INSUFFICIENT SOFT TISSUE TENSIONING; IT CAN HARDLY BE CAUSED BY STANDARD DEVICES. OTHER DEVICE INVOLVED IN THE EVENT: LINER: MPACT 01.32.3844MC DM CONVERTER LINER E/DMC LOT. 1903442 (K131458). BATCH REVIEW PERFORMED ON 20 NOVEMBER 2020: LOT 1903442: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-NOV-2019. EXPIRATION DATE: 2024-10-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

REVISION SURGERY FOR DISLOCATION OF THE DM POLY LINER FROM THE DM CONVERTER (DM POLY LINER AND COCR HEAD REMAINED COUPLED). THE SURGEON REVISED SUCCESSFULLY THE LINERS AND COCR HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1422643 LINER: MPACT DM DOUBLE MOBILITY HC LINER 22.2/DMC DUAL MOBILITY HC LINER MEH MEDACTA INTERNATIONAL SA 01.26.2246MHC 184721 07630030806834

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention