FDA Adverse Event Injury Summary report: N

TWIN-O-VAC

MDR report key: 109592 · Received July 25, 1997

Report

Report Number
109592
Event Type
Injury
Date Received
July 25, 1997
Date of Event
July 14, 1997
Report Date
July 24, 1997
Manufacturer
LABELED "CIG" AUSTRALIA
Product Code
FMT
Product Problem
Yes
Report Source
User Facility report
Reporter Location
WA, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

TWIN-O-VAC ATTACHED TO POLE OF INFANT WARMER IS USED EITHER TO DELIVER OXYGEN OR TO SUCTION. CAREGIVER, INTENDING TO SUCTION HOOKED TUBING UP TO OUTLET LABELED "OXYGEN" CAREGIVER UNABLE TO SEE OUTLET LABELED SUCTION DUE TO POSITION OF OUTLET AND POSITION OF LABELING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TWIN-O-VAC * FMT LABELED "CIG" AUSTRALIA * 521662

Patients

Seq Age Sex Outcome Treatment
1 0 DAY Hospitalization| L