FDA Adverse Event Malfunction Summary report: N

8015 ALARIS SYSTEM PC UNIT

MDR report key: 10956961 · Received December 7, 2020

Report

Report Number
2016493-2020-58470
Event Type
Malfunction
Date Received
December 7, 2020
Report Date
February 22, 2019
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801518
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACK WISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE SOURCE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

(B)(4). AFTER INVESTIGATION, THE DEVICE MODULE RUN TIME WAS CREATED BY COMPANY EMPLOYEES. THIS DEVICE MEETS THE CRITERIA PER (B)(4). INVESTIGATION SUMMARY: REPORT WON'T CONNECT TO ASM WAS CONFIRMED. FAULTY SIO BD IS THE MAIN CAUSE OF REPORT FAILURE. CONCLUSION: FAULTY SIO BD IS THE MAIN CAUSE OF REPORT FAILURE. REWORK: RECOMMEND REPLACING SIO BOARD (PART OR BOARD DESCRIPTION ASSY SIO BD PCU) PART NUMBER: TC10004047. DISPOSITION: ROUTE TO SERVICE FOR NORMAL PROCESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1426255 8015 ALARIS SYSTEM PC UNIT PUMP, INFUSION FRN CAREFUSION SD 8015 10885403801518

Patients

Seq Age Sex Outcome Treatment
1