FORCEPS/IRRIGATION PLUG (ISOLATED TYPE)
Report
- Report Number
- 8010047-2020-09909
- Event Type
- Malfunction
- Date Received
- December 7, 2020
- Report Date
- December 7, 2020
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- FGB
- PMA / PMN Number
- K912120
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC), BUT RETURNED TO OLYMPUS REPAIR CENTER FOR THE EVALUATION. THE EVALUATION OF THE DEVICE CONFIRMED THE FOLLOWINGS; COMPONENT ANALYSIS SHOWED THAT THE BROWN FOREIGN MATTER WAS RUST. OMSC GUESSED THAT THE SCREW COVER COMING OFF WAS CAUSED BY DETERIORATION OF THE ADHESIVE TO SECURE THE SCREW COVER DUE TO CHEMICAL OR PHYSICAL STRESS. THE RUST MAY HAVE BEEN CAUSED BY CORROSION OF RESIDUES SUCH AS CHEMICALS. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
A USER REPORTED THAT SOME THE PARTS OF THE DEVICE CAME OFF AND THE LEVER GOT LOOSE DURING REPROCESS. OLYMPUS INSPECTED THE DEVICE AT THE SERVICE DEPARTMENT OF OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) AND FOUND THAT THE SCREW COVER HAD COME OFF. IT WAS ALSO FOUND THAT BROWN FOREIGN MATTERS ADHERED BETWEEN THE STOP COCK AND THE MAIN BODY. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.