ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2020-57646
- Event Type
- Malfunction
- Date Received
- December 5, 2020
- Report Date
- November 12, 2020
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403810015
- PMA / PMN Number
- K133532
- Removal / Correction Number
- Z-2822-2020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE REPORTED ISSUE OF A DIM SEGMENT IS CONFIRMED BASED ON THE FIELD ACTION. NO FURTHER COMPLAINT INVESTIGATION IS NECESSARY AS CAPA 1143616 WAS OPENED TO ADDRESS THIS ISSUE. THE DEVICE IS USED FOR TREATMENT PURPOSES. CUSTOMERS RECEIVED NOTIFICATION OF THE FIELD ACTION. DEVICE REPAIR OR RETURNS ARE HANDLED WITHIN THE SCOPE OF THE FIELD ACTION. NO FURTHER INFORMATION WILL BE PROVIDED BY THE CUSTOMERS DUE TO THE FIELD ACTION.
IT WAS REPORTED THAT 9 LVP DEVICES HAD DIM SEGMENTS AND HAD TO BE REPLACED. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
NO DEVICE WILL BE RETURNED PER CUSTOMER. THE CUSTOMER COMPLAINT COULD NOT BE CONFIRMED BECAUSE THE DEVICE WAS NOT RETURNED FOR FAILURE INVESTIGATION. THE ROOT CAUSE OF THIS FAILURE WAS NOT IDENTIFIED.
IT WAS REPORTED THAT 9 LVP DEVICES HAD DIM SEGMENTS AND HAD TO BE REPLACED. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1418750 | ALARIS PUMP MODULE | PUMP, INFUSION | FRN | CAREFUSION SD | 8100 | 10885403810015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |