FDA Adverse Event Malfunction Summary report: N

TAYLOR-WHARTON

MDR report key: 109526 · Received July 3, 1997

Report

Report Number
109526
Event Type
Malfunction
Date Received
July 3, 1997
Date of Event
January 1, 1995
Report Date
May 30, 1997
Manufacturer
TAYLOR-WHARTON
Product Code
JRM
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

SOMETIME IN AUGUST/SEPTEMBER 1995, CONDENSATION BUILT UP INSIDE THE FREEZER CAUSING THE LID TO BE FROZEN IN PLACE ABOVE THE NORMAL POSITION. CONSEQUENTLY THE LID WAS REPLACED. IN MARCH 1997, THE LID REPLACEMENT BEGAN TO SHOW SIGNS OF LIFTING WITH ACCUMULATED ICE BUILD-UP. A NEW FREEZER WAS ORDERED AND TRANSFER OF BONE MARROW PRODUCT INITIATED ON 5-9-97. DURING THE PROCESS OF TRANSFERRING PRODUCT TO THE NEW FREEZER IT WAS NOTED THAT SOME OF THE PRODUCT BAGS HAD LEAKED AND NOW CONTAINED LIQUID NITROGEN. DURING MONTH OF APRIL 1997, FREEZER ALARM WOULD GO OFF INDICATING NEED FOR MORE LIQUID NITROGEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAYLOR-WHARTON LIQUID NITROGEN STORAGE FREEZER JRM TAYLOR-WHARTON K SERIES *

Patients

Seq Age Sex Outcome Treatment
1 * Other GAMBRO BAGS