FDA Adverse Event Injury Summary report: N

EZTETIC DENTAL IMPLANT, 3.1MM DIAMETER, 2.9MMD PLATFORM, 13MM LENGTH

MDR report key: 10951027 · Received December 5, 2020

Report

Report Number
0002023141-2020-02120
Event Type
Injury
Date Received
December 5, 2020
Date of Event
September 16, 2020
Report Date
February 11, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K142082
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: METHOD CODE WAS ADDED: 3331 AND 4109. H6: RESULTS CODE WAS ADDED: 213. H6: CONCLUSIONS CODE WAS ADDED: 67. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOTBE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1230315). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD (OP#140) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1230315) FOR SIMILAR EVENT AND ONE (1) OTHER RELEVANT COMPLAINT WAS IDENTIFIED, ((B)(4)). INVESTIGATION HAS IDENTIFIED THAT THE MOST LIKELY PROBABLE CAUSES ARE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION HAS BEEN RECEIVED.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE # 11 DEVELOPED AN INFECTION AND WAS REMOVED. FISTULA PRESENT MULTIPLE TIMES OVER THE 3 MONTH PERIOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1415467 EZTETIC DENTAL IMPLANT, 3.1MM DIAMETER, 2.9MMD PLATFORM, 13MM LENGTH DENTAL IMPLANT DZE ZIMMER DENTAL 1230315

Patients

Seq Age Sex Outcome Treatment
1 18 YR Required Intervention