DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2020-146379
- Event Type
- Malfunction
- Date Received
- December 4, 2020
- Date of Event
- November 7, 2020
- Report Date
- February 10, 2021
- Manufacturer
- DEXCOM, INC.
- Product Code
- QBJ
- UDI-DI
- 00386270000385
- PMA / PMN Number
- DEN170088
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). B5: DESCRIBE EVENT OR PROBLEM - ADDITIONAL . H2: ADDITIONAL INFORMATION/DEVICE EVALUATION . H3: DEVICE EVALUATED BY MANUFACTURER - ADDITIONAL . H6: ADVERSE EVENT PROBLEM - ADDITIONIAL - CORRECTION PLEASE OMIT H6 CODES FROM INITIAL SUBMISSION 11, 3233.
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0 VDC. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
(B)(4).
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1412116 | DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | QBJ | DEXCOM, INC. | 9445-02 | 5275851 | 00386270000385 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |