FDA Adverse Event Malfunction Summary report: N

EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

MDR report key: 10944729 · Received December 4, 2020

Report

Report Number
8010047-2020-09810
Event Type
Malfunction
Date Received
December 4, 2020
Report Date
February 23, 2021
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
ODG
UDI-DI
04953170355479
PMA / PMN Number
K093395
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE PROBABLE CAUSE IS LIKELY EXTERNAL FORCE WAS APPLIED CREATING CHIPPING OF THE ADHESIVE AT THE DISTAL END. THIS ISSUE IS ADDRESSED IN THE INSTRUCTIONS FOR USE (IFU): "·IMPORTANT INFORMATION ¿ PLEASE READ BEFORE USE DO NOT APPLY SHOCK TO THE DISTAL END OF THE INSERTION SECTION, PARTICULARLY THE ULTRASOUND TRANSDUCER AND THE OBJECTIVE LENS SURFACE AT THE DISTAL END. VISUAL ABNORMALITIES MAY RESULT.".

Additional Manufacturer Narrative · 1

THE DEVICE REFERENCED IN THIS REPORT WAS RETURNED TO OLYMPUS FOR PHYSICAL EVALUATION. THE INVESTIGATION IS ONGOING. PHYSICAL EVALUATION OF THE RETURNED DEVICE REVEALS THE FOLLOWING: THE UM CONNECTOR IS NOT ATTACHED, THEREFORE LEAK TEST CAN NOT BE PERFORMED. UM IMAGE WAS NOT DETECTABLE DUE TO MULTIPLE BROKEN ELEMENTS. THE FORCEPS COVER GLUE WAS PEELING. THIS REPORT WILL BE UPDATED UPON COMPLETION OF THE INVESTIGATION OR UPON RECEIPT OF ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT IS REPORTED WHILE PREPARING A IS EXERA II ULTRASOUND GASTROVIDEOSCOPE FOR USE, SOMEONE TRIPPED OVER THE DEVICE AND THE U-CONNECTOR WAS COMPLETELY BROKEN OFF WITH WIRES HANGING OUT. THERE WAS NO PATIENT CONTACT/IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1409132 EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE ULTRASOUND GASTROVIDEOSCOPE ODG OLYMPUS MEDICAL SYSTEMS CORP. GF-UCT180 04953170355479

Patients

Seq Age Sex Outcome Treatment
1