LUXSTAR STANDARD POST-MOUNT LIGHT
Report
- Report Number
- 1017522-2020-00014
- Event Type
- Malfunction
- Date Received
- December 4, 2020
- Date of Event
- October 28, 2020
- Report Date
- December 3, 2020
- Manufacturer
- KAVO DENTAL TECHNOLOGIES, LLC
- Product Code
- EAZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT HAS A THREE PRONG LOCKING MECHANISM WHICH LOCKS THE LENS HEAT SHIELD/COVER INTO PLACE. THE LENS HEAT SHIELD/COVER HAS TO BE REMOVED BY THE END USER WHEN CLEANING THE LENS OR REPLACING THE HALOGEN BULB DURING ROUTINE OPERATOR MAINTENANCE. THE INSTALLATION INSTRUCTIONS, USE AND CARE INSTRUCTIONS, AND LABELING CLEARLY STATE TO ENSURE THE LENS HEAT SHIELD/HOLDER IS PROPERLY SECURE BY SNAPPING THE PART BACK INTO PLACE. THE LIGHT WAS MANUFACTURED IN OCTOBER 2001 AND IS OVER 19 YEARS OLD AND PAST ITS EXPECTED LIFE. THIS CONCLUDES THE INVESTIGATION.
A DENTAL PROFESSIONAL WAS TREATING A PATIENT WHEN THE LENS HEAT SHIELD COVER FELL OFF A MARUS DENTAL LIGHT HITTING THE PATIENT CAUSING A SMALL BURN. THERE WERE NO SERIOUS INJURIES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1412414 | LUXSTAR STANDARD POST-MOUNT LIGHT | DENTAL LIGHT | EAZ | KAVO DENTAL TECHNOLOGIES, LLC | UL1000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |