FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 10939102 · Received December 3, 2020

Report

Report Number
1911916-2020-01091
Event Type
Malfunction
Date Received
December 3, 2020
Date of Event
November 6, 2020
Report Date
November 17, 2020
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903096535
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED. (B)(4). INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 309653 AND LOT NUMBER 0140473. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. INVESTIGATION CONCLUSION: COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE DESCRIPTION: BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. RATIONALE: FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 50ML LL TIP 1ML WAS MISSING SCALE MARKINGS ON 25 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 309653 BATCH NO. 0140473. IT WAS REPORTED THAT SOME SYRINGES DO NOT HAVE MARKINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1408083 SYRINGE 50ML LL TIP 1ML PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 309653 0140473 30382903096535

Patients

Seq Age Sex Outcome Treatment
1