FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 30GA 8MM 7BAG 420CAS JP

MDR report key: 10936625 · Received December 2, 2020

Report

Report Number
1920898-2020-01661
Event Type
Malfunction
Date Received
December 2, 2020
Date of Event
November 9, 2020
Report Date
April 9, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY CUSTOMER RETURNED PHOTOS OF A 3/10CC, 8MM, 30G SYRINGE IN A POLY BAG FROM LOT # 9210505. CUSTOMER STATES THAT THERE IS NEEDLE/SHIELD SEPARATION FROM THE BARREL. THE PHOTOS WERE EXAMINED AND EXHIBITED THE HUB-NEEDLE/SHIELD ASSEMBLY SEPARATED FROM THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 9210505 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA1630423 HAS BEEN OPENED TO ADDRESS THIS ISSUE. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 0.3ML 30GA 8MM 7BAG 420CAS JP SEPARATED FROM NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER REPORTED ABOUT NEEDLE/SHIED SEPARATION FROM THE BARREL.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 0.3ML 30GA 8MM 7BAG 420CAS JP SEPARATED FROM NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER REPORTED ABOUT NEEDLE/SHIED SEPARATION FROM THE BARREL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1402357 SYRINGE 0.3ML 30GA 8MM 7BAG 420CAS JP SYRINGE FMF BD MEDICAL - DIABETES CARE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1