FDA Adverse Event Malfunction Summary report: N

ACUSNARE POLYPECTOMY SNARE

MDR report key: 10935020 · Received December 2, 2020

Report

Report Number
1037905-2020-00519
Event Type
Malfunction
Date Received
December 2, 2020
Date of Event
November 5, 2020
Report Date
December 2, 2020
Manufacturer
COOK ENDOSCOPY
Product Code
FDI
UDI-DI
00827002226470
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONTINUED: INITIAL REPORTER; OCCUPATION: OTHER-HEALTHCARE PROFESSIONAL. CONTINUED: PMA/510(K) # K191048. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. THE DEVICE HISTORY RECORD CONTAINS NONCONFORMANCES THAT COULD POTENTIALLY BE RELATED TO THIS REPORT. THE DEVICE GOES THROUGH DIFFERENT INSPECTIONS PRIOR TO LEAVING THE FACILITY. THESE INSPECTIONS WOULD HAVE REMOVED ANY NONCONFORMING DEVICES PRIOR TO DISTRIBUTION. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. THE INSTRUCTIONS FOR USE CONTAINS THE FOLLOWING INFORMATION TO ASSIST WITH PROPER SETUP AND USE OF THE DEVICE: "FULLY RETRACT AND EXTEND SNARE TO CONFIRM SMOOTH OPERATION OF DEVICE." PRIOR TO DISTRIBUTION, ALL ACUSNARE POLYPECTOMY SNARES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ENDOSCOPIC POLYPECTOMY, THE PHYSICIAN USED A COOK ACUSNARE POLYPECTOMY SNARE AND THERE WAS RESISTANCE IN OPENING AND CLOSING THE DEVICE [DIFFICULTY WITH RETRACTION SUBJECT OF REPORT]. A NEW DEVICE WAS OPENED AND USED TO COMPLETE THE INTENDED PROCEDURE SUCCESSFULLY." A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1398420 ACUSNARE POLYPECTOMY SNARE FDI, SNARE, FLEXIBLE FDI COOK ENDOSCOPY G22647 W4354297 00827002226470

Patients

Seq Age Sex Outcome Treatment
1