COCR BALL HEAD 12/14 32 SIZE S -3.5
Report
- Report Number
- 3005180920-2020-00872
- Event Type
- Injury
- Date Received
- December 1, 2020
- Date of Event
- November 8, 2020
- Report Date
- December 1, 2020
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JDI
- UDI-DI
- 07630030805158
- PMA / PMN Number
- K072857
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 20 NOVEMBER 2020: LOT 188824: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 7-MAR-2019. EXPIRATION DATE: 2024-02-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. OTHER DEVICE INVOLVED IN THE EVENT: LINER: MPACT 01.32.3241HCT FLAT PE HC LINER ¿32/D LOT. 1902883 (K103721), BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 20 NOVEMBER 2020: LOT 1902883: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 8-MAY-2019. EXPIRATION DATE: 2024-04-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
THE PATIENT CAME IN REPORTING PAIN DUE TO A DISLOCATION OF THE HEAD FROM THE LINER 4 DAYS AFTER THE PRIMARY SURGERY. THE CAUSE OF THE DISLOCATION IS UNKNOWN. THE SURGEON REVISED THE HEAD AND LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1396184 | COCR BALL HEAD 12/14 32 SIZE S -3.5 | BALL HEAD | JDI | MEDACTA INTERNATIONAL SA | 01.25.021 | 188824 | 07630030805158 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |