MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2020-15240
- Event Type
- Injury
- Date Received
- December 1, 2020
- Date of Event
- October 22, 2020
- Report Date
- November 6, 2020
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317003854
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
ON DECEMBER 2, 2020, MENTOR BECAME AWARE THE PATIENT WOULD UNDERGO EXPLANTATION ON (B)(6) 2020. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON (B)(6) 2021, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON (B)(6) 2021, MENTOR COMPLETED EVALUATION OF THE DEVICE. DEVICE EVALUATION SUMMARY: DURING THE VISUAL INSPECTION, THE DEVICE WAS FOUND TO BE RUPTURED AND INCOMPLETE. ADDITIONALLY, AN ANOMALY WAS OBSERVED ON THE EDGES OF THE RUPTURE CONSISTENT WITH SHELL ABRASION. THE EVALUATION DETERMINED THAT THE LOSS OF SHELL INTEGRITY IS CONSISTENT WITH A RUPTURE CAUSED BY WEAR. SHELL ABRASION SUGGESTS IN-VIVO FOLDING OR CREASING OF THE DEVICE. THIS MAY BE THE RESULT OF THE FOLLOWING FACTORS: CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE SUCH AS TOO SMALL BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. IN SOME CASES, THE BREAST IMPLANTS MAY ALSO WEAR OUT OVER TIME. A SECOND PRODUCT WAS RECEIVED (PRODUCT CODE 3507501BC AND LOT NUMBER 5963498). NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. RUPTURE COMPLAINT INFORMATION IS CONSISTENTLY ANALYZED AND MONITORED BY MENTOR QUALITY ASSURANCE TO DETERMINE WHEN FURTHER ACTION IS NECESSARY. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE HAS BEEN RECEIVED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6) YEAR-OLD FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 750CC MENTOR MEMORYGEL BREAST IMPLANT AND EXPERIENCED A RUPTURE ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED BY A PHYSICIAN. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1395974 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 3507501BC | 5993161 | 00081317003854 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Other |