FDA Adverse Event Malfunction Summary report: N

ACUSNARE POLYPECTOMY SNARE

MDR report key: 10920918 · Received November 30, 2020

Report

Report Number
1037905-2020-00514
Event Type
Malfunction
Date Received
November 30, 2020
Date of Event
October 29, 2020
Report Date
November 30, 2020
Manufacturer
COOK ENDOSCOPY
Product Code
FDI
UDI-DI
00827002226470
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

PMA/510(K) #: K191048. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING INFORMATION TO ASSIST WITH PROPER SETUP AND USE OF THE DEVICE: "BEFORE USING THIS DEVICE, FOLLOW RECOMMENDATIONS PROVIDED BY THE ELECTROSURGICAL UNIT MANUFACTURER TO ENSURE PATIENT SAFETY THROUGH PROPER SELECTION, PLACEMENT AND UTILIZATION OF THE PATIENT RETURN ELECTRODE. ENSURE THAT A PROPER PATH FROM THE PATIENT RETURN ELECTRODE TO THE ELECTROSURGICAL UNIT IS MAINTAINED THROUGHOUT THE PROCEDURE." PRIOR TO DISTRIBUTION, ALL ACUSNARE POLYPECTOMY SNARES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED AND THIS REPRESENTS AN ISOLATED OCCURRENCE. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ENDOSCOPIC POLYPECTOMY, THE PHYSICIAN USED A COOK ACUSNARE POLYPECTOMY SNARE. WHEN THEY USED THE DEVICE WITH CAUTERY, IT WOULD NOT CUT THE TISSUE UNTIL THEY CHANGED THE SETTINGS TO MORE OF A "COAGULATION." THE SNARE DID NOT CUT ON THE "BLEND" SETTING BUT CUT ON THE "COAG." UNKNOWN IF THE GROUNDING PAD MAY HAVE BEEN IN THE WRONG POSITION. DEVICE DID WORK WELL ENOUGH TO COMPLETE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1388024 ACUSNARE POLYPECTOMY SNARE FDI, SNARE, FLEXIBLE FDI COOK ENDOSCOPY G22647 W4343222 00827002226470

Patients

Seq Age Sex Outcome Treatment
1