8015 ALARIS SYSTEM PC UNIT
Report
- Report Number
- 2016493-2020-52056
- Event Type
- Malfunction
- Date Received
- November 30, 2020
- Report Date
- December 11, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403801518
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). 8015 100.5010 ERROR. POLO CORRUPT. BIOMED JUST REPLACED THE LOGIC BOARD AND GETTING THE 100.5010 ERROR. RECOMMEND HE PURCHASE COMPACT FLASH CARDS SO HE CAN FLASH THE 8015 UNIT. GAVE HIM PART NUMBER TO COMPACT FLASH CARDS AND TRANSFER TO DAVID COM FOR PRICING. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1382127 | 8015 ALARIS SYSTEM PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 | 10885403801518 |