FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 3ML LL

MDR report key: 10910018 · Received November 26, 2020

Report

Report Number
1213809-2020-00845
Event Type
Malfunction
Date Received
November 26, 2020
Date of Event
November 2, 2020
Report Date
December 2, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096573
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVALUATION? YES. D.10. RETURNED TO MANUFACTURER ON: (B)(6)2020. H.6. INVESTIGATION: FOUR PHOTOS DISPLAYING A TOTAL OF THIRTEEN OF 3ML SYRINGES IN FULLY SEALED BLISTER PACKS (P/N 309657) WERE RECEIVED AND EVALUATED. IT WAS OBSERVED ONE OF THE SYRINGES HAD THE BATCH NUMBER 5300809 EMBOSSED ON THE PEEL TAB. TWELVE OF THE PACKAGES HAD THE BATCH NUMBER 6110527 PRINTED ON THE PEEL TAB. NO DEFECTS WERE OBSERVED. BATCH 5300809 WAS MANUFACTURED IN 2015, PRIOR TO THE UDI REQUIREMENTS THAT MANDATED MARKING INDIVIDUAL PACKAGING WITH EXPIRATION DATE. BATCH 5300809 WAS MANUFACTURED CORRECTLY PER PRODUCT DESIGN AT THAT TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE LABEL WAS MISSING THE LOT AND AND EXPIRATION DATE WITH A BD SYRINGE 3ML LL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT EXPIRATION AND LOT NUMBER WERE MISSING.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LABEL WAS MISSING THE LOT, AND EXPIRATION DATE WITH A BD SYRINGE 3ML LL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT EXPIRATION AND LOT NUMBER WERE MISSING.