T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2020-132544
- Event Type
- Malfunction
- Date Received
- November 25, 2020
- Date of Event
- November 9, 2020
- Report Date
- November 25, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613380
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE PUMP USER GUIDE STATES: WARNING - NEVER FILL YOUR TUBING WHILE YOUR INFUSION SET IS CONNECTED TO YOUR BODY. ALWAYS ENSURE THAT THE INFUSION SET IS DISCONNECTED FROM YOUR BODY BEFORE FILLING THE TUBING. FAILURE TO DISCONNECT YOUR INFUSION SET FROM YOUR BODY BEFORE FILLING THE TUBING CAN RESULT IN OVER DELIVERY OF INSULIN. THIS CAN CAUSE SERIOUS INJURY OR DEATH FROM VERY LOW BLOOD GLUCOSE.
IT WAS REPORTED THAT THE CUSTOMER INADVERTENTLY PERFORMED THE LOAD FILL TUBING SEQUENCE WHILE CONNECTED TO THE SITE. THE CUSTOMER'S BLOOD GLUCOSE RANGED FROM 120-387 MG/DL. TANDEM TECHNICAL SUPPORT EDUCATED THE CUSTOMER TO NOT BE CONNECTED TO THE PUMP DURING THE FILL TUBING PROCESS. REPORTEDLY, CUSTOMER SUCCESSFULLY COMPLETED LOAD SEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1371770 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613380 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |