FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 10906692 · Received November 25, 2020

Report

Report Number
3013756811-2020-132544
Event Type
Malfunction
Date Received
November 25, 2020
Date of Event
November 9, 2020
Report Date
November 25, 2020
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613380
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE PUMP USER GUIDE STATES: WARNING - NEVER FILL YOUR TUBING WHILE YOUR INFUSION SET IS CONNECTED TO YOUR BODY. ALWAYS ENSURE THAT THE INFUSION SET IS DISCONNECTED FROM YOUR BODY BEFORE FILLING THE TUBING. FAILURE TO DISCONNECT YOUR INFUSION SET FROM YOUR BODY BEFORE FILLING THE TUBING CAN RESULT IN OVER DELIVERY OF INSULIN. THIS CAN CAUSE SERIOUS INJURY OR DEATH FROM VERY LOW BLOOD GLUCOSE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER INADVERTENTLY PERFORMED THE LOAD FILL TUBING SEQUENCE WHILE CONNECTED TO THE SITE. THE CUSTOMER'S BLOOD GLUCOSE RANGED FROM 120-387 MG/DL. TANDEM TECHNICAL SUPPORT EDUCATED THE CUSTOMER TO NOT BE CONNECTED TO THE PUMP DURING THE FILL TUBING PROCESS. REPORTEDLY, CUSTOMER SUCCESSFULLY COMPLETED LOAD SEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1371770 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 00850006613380

Patients

Seq Age Sex Outcome Treatment
1 63 YR