FDA Adverse Event Malfunction Summary report: N

SLIDING MECHANISM

MDR report key: 10906665 · Received November 25, 2020

Report

Report Number
2939274-2020-05324
Event Type
Malfunction
Date Received
November 25, 2020
Date of Event
October 30, 2020
Report Date
October 30, 2020
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTD
UDI-DI
10886982188779
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. DEVICE HISTORY LOT PART: 314.291, LOT: 170527-902, MANUFACTURING SITE: SELZACH, SUPPLIER: (B)(4). RELEASE TO WAREHOUSE DATE: APRIL 30, 2018 . A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE USING PROXIMAL FEMORAL NAILING SYSTEM (TFNA). DURING THE PROCEDURE, THE SLIDING MECHANISM CLAMP FELL OFF THE PATIENT AND HIT THE GROUND, BREAKING THE HANDLE IN HALF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNKNOWN NUMBER OF SURGICAL DELAY. THE PATIENT WAS STABLE. THIS REPORT IS FOR ONE (1) SLIDING MECHANISM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1371664 SLIDING MECHANISM FORCEPS HTD WRIGHTS LANE SYNTHES USA PRODUCTS LLC 314.291 170527-902 10886982188779

Patients

Seq Age Sex Outcome Treatment
1 79 YR