HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2020-00658
- Event Type
- Malfunction
- Date Received
- November 25, 2020
- Report Date
- December 18, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT THE DEVICE WAS PLACED INSIDE THE OPERATING THEATRE DURING USE WITH THE FAN DIRECTED OPPOSITE TO THE PATIENT AT AN ESTIMATED DISTANCE OF TWO (2) METERS FROM THE SURGERY FIELD. IN ADDITION, LIVANOVA LEARNED THAT THE DEVICE IS CLEANED REGULARLY IN TERMS OF TIMELINES AND THAT THE HOSPITAL USES A RECIPE SIMILAR TO CLOROX DISINFECTANT. THIS IS NOT IN ACCORDANCE WITH THE LIST OF APPROVED DISINFECTANTS PRESCRIBED WITHIN THE IFU. THIS MAY HAVE LED/CONTRIBUTED TO THE REPORTED CONTAMINATION.
SEE INITIAL REPORT.
THERE WAS NO KNOWN PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA DEUTSCHLAND IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015 H10: LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN ROME, ITALY. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DEVICE WAS FOUND TO BE CONTAMINATED WITH MYCOBACTERIUM CHIMAERA. THE LABORATORY REPORT CONFIRMING THE REPORTED CONTAMINATION HAS BEEN PROVIDED TO LIVANOVA AND IT REFERS TO PRE-DISINFECTION CYCLE SAMPLES. THERE IS NO KNOWN PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1369175 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |