8100 ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2020-48574
- Event Type
- Malfunction
- Date Received
- November 24, 2020
- Report Date
- September 19, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403810015
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). PLEASE REFER TO FILE #2018-033988-265859 FROM CAD. (B)(4). NPI SN (B)(4). (B)(6) HAD AN ERROR CODE 210.6020 ON LVP8100. THE UNIT STILL UNDER WARRANTY. (B)(6) WILL SEND IT IN FOR WARRANTY REPAIR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1357586 | 8100 ALARIS PUMP MODULE | PUMP, INFUSION | FRN | CAREFUSION SD | 8100 | 10885403810015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |